官方网站

XT5418-DPC series photostability test chambers for

Function introduction:



Available with capacity of 275L, 480L, 800L and 1200L for varieties of applications 

Guidelines for temperature testing, humidity testing, accelerated testing, long term testing and, visual and UVA exposure testing

Patented stepper valve hot gas by-passing and dynamic temperature and humidity control for stability up to ±0.1 °C and ±2.0 %RH, for continuous reliable operation

1.Available with capacity of 275L, 480L, 800L and 1200L for varieties of applications 

2.Full range of features to support ICH Q1A (R2) and Q1 B 

3.Guidelines for temperature testing, humidity testing, accelerated testing, long term testing and,

 visual and UVA exposure testing

4.Patented stepper valve hot gas by-passing and dynamic temperature and humidity control for stability 

up to ±0.1 °C and ±2.0 %RH, for continuous reliable operation

5..7 in high resolution true color TFT touch screen for data input for equipment opera-tion, programming 

of temperature and humidity, alarm history and operating status, in text,graphic symbols and curves .

6.Optional data acquisition and recording system meeting requirements for FDA, CFR  21 Part 11Replacement

 for Binder KBF & KBF P, Memmert ICH L and MMM Climacell

7.Specifications:

                 Models

  XT5418-DPC275

  XT5418-DPC480

  XT5418-DPC800

XT5418-DPC1200

Effective capacity  L

          275

         480

         800

       1200

Temperature range@20℃

      +5~+65

      +5~+65

     +5~+65

     +5~+65

Temperature stability ±℃

       0.1~0.2

      0.1~0.2

      0.1~0.2

      0.1~0.2

Temperature deviation  ℃

          1.0

          1.0

         1.0

         1.0

Humidity range(std) %RH

       40~95

        40~95

      40~95

       40~95

Humidity range(opt) %RH

       20~95

        20~95

      20~95

       20~95

Humidity stability±%RH

        1~2

         1~2

         1~2

        1~2

Humidity deviation  %RH

        3~5

         3~5

         3~5

        3~5

Heater wattage W

          450

          450

          950

        950

Cooling capacity@20℃ W

          300

          400

          650

        950

Interior size(WDH)  mm

 535×480×950

635×560×1350

935×640×1350

1235×720×1350

Overall size (WDH) mm

640×660×1600

740×740×1972

1040×820×1972

1340×900×1972

Air movement

  Horizontal

   Horizontal

   Horizontal

    Horizontal

VIS canopies&shelves (std/max)EA

        1/3

        1/4

          1/4

         1/4

UVA canopies&shelves (std/max)EA

        1/3

        1/4

          1/4

         1/4

Illumination intensity,FLCW,D65 Lux

        8000

        8000

         8000

         8000

Illumination intensity,UVA mW/m2

       2000

        2000

          2000

         2000