XT5618D12/XT5618D12HT series Refrigerated and heating circulators
- Function introduction:
Patented stepper valve hot gas by-passing cascaded refrigeration for dynamic cooling capacity control and ACC temperature up to 200℃ ,
Choose from models for minimum temperatures down to -60℃,-70℃,-80℃ and -90℃ ,
Powerful multi-stage immersion pump for maximum pressure up to 2 bar and flow rate up to 30L/min
1、Patented stepper valve hot gas by-passing cascaded refrigeration for dynamic cooling capacity control
and ACC temperature up to 200℃;
2、Choose from models for minimum temperatures down to -60℃,-70℃,-80℃ and -90℃;
3、Powerful multi-stage immersion pump for maximum pressure up to 2 bar and flow rate up to 30L/min
Completed with adjustable overheating protection and additional low level cut out for flammable liquid
heat transfer media
4、Completed features to meet requirements of IEC 61010-1 and IEC 61010-2-012 for safety
Alternative forpower supplies for Europe,North America, South America and Japan
5、Specifications:
Models |
XT5618D12 –R60HG |
XT5618D12 –R70HG |
XT5618D12 –R80HG |
XT5618D12 –R90HG |
Temp range@20 D12 |
-60~+95 |
-70~+95 |
-80~+95 |
-88~+95 |
Temp range@20 D12HT |
-60~+200 |
-70~+200 |
-80~+200 |
-88~+200 |
Temperature stability ±℃ |
0.05~0.1 |
0.05~0.1 |
0.05~0.1 |
0.05~0.1 |
Heater wattage |
1800W |
1800W |
1800W |
1800W |
Cooling capacity@20℃ |
900W |
1250W |
1200W |
1200W |
Cooling capacity@0℃ |
800W |
1100W |
1000W |
1000W |
Cooling capacity@ -20℃ |
400W |
1000W |
500W |
500W |
Cooling capacity@ -70℃ |
- |
450W |
120W |
120W |
Pump pressure max |
1.8~2.0bar |
1.4~1.6bar |
1.0~1.2bar |
1.0~1.2bar |
Pump flow-rate max |
30L/min |
30L/min |
30L/min |
30L/min |
Interior size mm |
215W×350D×200H |
215W×350D×200H |
215W×350D×200H |
215W×350D×200H |
Bath opening mm |
215W×150D |
215W×150D |
215W×150D |
215W×150D |
Overall size mm |
475W×575D×915H |
475W×575D×915H |
475W×575D×915H |
475W×575D×915H |